Verification caliber |
DN15-50 |
Capacity |
DN15 10pcs、DN20 8pcs、DN25 6pcs、
DN32 2pcs、DN40 2pcs、DN50 1pcs
|
Measurement Range |
(0.06~40)m³/h |
Electronic Scale |
①Max=30kg,6000e Max=600kg,3000e |
Device Expansion Uncertainty |
Master meter method: < 0.5 %
Start-stop weighing method: < 0.2 % (k = 2)
|
Heat Meter Type |
Ultrasonic heat meter
Heat meter
|
constant temperature bath |
Integrated high/low temperature bath
Measuring range: 0 °C – 100 °C
Temperature fluctuation: ±0.005 °C per 10 min
|
constant - temperature standard |
Standard platinum resistance thermometer (SPRT)
constant temperature bath
|
Temperature difference |
Horizontal temperature difference: ≤ 0.006 °C
Vertical temperature difference: ≤ 0.006 °C
|
Material of the device |
Made of high-quality industrial aluminum and stainless steel 304,durable and excellent appearance
|
Temperature control range |
Ambient~55 °C;
heat-up time for the heat meter
from 25°C to 55°C: 4 h
|