| Device Uncertainty |
Urel≤0.2%(k=2) |
| Flow Stability |
1.0% for start‑stop method; 0.2% for other methods |
| Test Medium |
Purified tap water |
| Operating Humidity |
≤ 93% RH |
| Operating Water Temperature |
20 ℃ ± 10 ℃ |
| Water Temperature Fluctuation in Single Measurement |
Water temperature fluctuation in a single measurement |
| Water Fluctuation Coefficient per Cycle |
≤ ± 0.1% |
| Water Source Working Pressure |
0.0–0.5 MPa, adjustable |
| Air Source Working Pressure |
0.4–0.7 MPa |
| Device Operating Temperature |
4–50 ℃ |
| Number of Meters Under Test |
DN15: 10 pcs | DN20: 8 pcs | DN25: 6 pcs | DN40 / DN50 / DN65: 5 pcs | DN80 / DN100: 3 pcs |
| Water Pump |
1 pc: H=20 m, Q=2 m³/h, P=0.5 kW; 1 pc: H=30 m, Q=40 m³/h, P=7.5 kW; 1 pc: H=30 m, Q=180 m³/h, P=22 kW |
| Water Tank |
15000 L |
| Commutator |
DN25 & DN100 cross‑fork pneumatic open commutator, low noise and high sensitivity; travel difference < 10 ms; uncertainty within 0.002% |
| Meter Clamping Mechanism |
Dual‑cylinder outward‑push series pneumatic clamping mechanism; single pneumatic clamps for front 10 and rear 5 positions |
| Water Meter Seal Test |
Pressure range: (0–2.0) MPa |
| Bore Diameter |
Φ50 mm, telescopic stroke ≥ 80 mm |
| Timer |
(10–100) kHz (adjustable); better than 10⁻⁶ |
| Calibration Signal |
Pulse signal, f = (0–10) kHz |
| Vacuum System |
Differential pressure vacuum generator for vacuum water filling to keep water meters fully filled |
| Supported Operation Modes |
Fully automatic, semi‑automatic, touchscreen manual operation |
| Acquisition Equipment |
5 sets of mobile phones with 50 MP resolution, 8G / 128G storage |
| Standard Device |
Volumetric method |
| Flow Range |
(0.010-130)m³/h |
| Calibration Method |
Start‑stop method |
Diversion method |
QTY |
Rotameter Instantaneous Flow Display Basic |
LZ/W30-25,9~160L/H |
1 |
| LZ/W30-25,20~500L/H |
1 |
| LZ/W30-25 120-1600L/H |
1 |
| LZ/W30-50 500-8000L/H |
1 |
Electromagnetic Flow Meter Instantaneous Flow Display Advanced (Class 0.5) |
DN2:3-90L/H |
1 |
| DN10:20-500L/H |
1 |
| DN25:500-800L/H |
1 |
| DN50:8000-40000L/H |
1 |
| DN100 10-200m³/H |
1 |
Volumetric Vessel Volumetric Method Standard |
10L-20L |
1 |
| 50L-105L |
1 |
| 300L-500L |
1 |
| 1500L-3000L |
1 |